- Description
- Reviews (0)
- TECHNICAL SPECIFICATIONS
Description
Automatic analyzer random access for clinical-chemistry and immunoturbidimetry computer controlled. Automatic – upon programmation – pathological results repetition with or without automatic dilution. Automatic – upon programmation – pathological results in the panic range repetition. Urgent tests in actual time (STATs).
Sophisticated software for the system control. Software running under Windows 2000 Professional®. 40 on line tests mono or double reagent, refrigerated. 78 positions sample tray, programmable for samples, standards, controls and STATS.
Analyzer throughput: 360 test/h without ISE module. ISE module: 240 test/h. Two bar code readers for the positive identification of samples and reagents. Easy maintenance. No consumables.
Automatic washing of the reading cuvettes. Automatic – upon programmation – calibration of the I.S.E. module. Relation tests. Real pre-dilution up to 1:500. Clot Sensor: For Samples and Reagents.
Log file for data and performed operations.Vocal and on screen error messages.Touch screen monitor and wireless keyboard and mouse.
Be the first to review “BT 3500”
TECHNICAL SPECIFICATIONS | |
Type | Stand Alone, Multiparametric System |
Analysis Type | Clinical Chemistry, Turbidimetric, ISE |
Throughput | Up to 360 Chem. Tests/hr + up to 240 ISE Tests/hr |
ISE Module | Na, K, Cl, CO² (indirect potentiometry) |
Reagent Tray Positions | Up to 80 (40 large bottles + 40 small bottles) |
Reagent Tray Refrigeration | From 8°C to 15°C (programmable) |
Sample Tray Positions | 78 (62 Samples + 16 Standard and Controls) |
Sample tubes | Primary tubes and sample cups |
Reaction volume | 180 up to 400 µl |
Sample volume | 1.8 µl up to 100 µl |
Reading cuvettes | 45 non-disposable optical glass |
Sampling Arm | 2 for Serum and Reagents |
Number of position | 78 Total – 52 Samples – 26 Standards and Control |
Test Modes | Random Access, Batch, STAT |
Test Directory | Up to 500 |
Temperature control | Peltier System OFF/30/32/37°C (±0,2°C) |
Vacuum Pump | External |
Photometer | Solid state 10 channel + Reference channel |
Photometer Accuracy | ± 1% (0-2 O.D.), + 2,5% (2-3 O.D.) |
Photometer Sensitivity | ± 0,001 Abs |
Diluter | 1 – Long life ceramic 340 µl |
Diluter Accuracy | ± 0,1% f.s. |
Diluter Sensitivity | ± 0.1 µl |
Clot Sensor | For Sample and Reagents |
Bar Code | For Sample and Reagent |
Remote Diagnostics | Via Modem (optional) |
Computer | CPU Pentium M® 1.6Ghz or more, Monitor color LCD 12″, DVD Rom, HD=>80 Gb |
Touch-screen | Yes: Glass Protected resistive kind |
Keyboard | Standard IBM® Compatible – wireless |
Mouse | Standard – wireless |
Software | Windows 2000® PRO |
Process electronic | µP. 1²CBUS Protocol |
Interface | RS232, Bidirectional + USB + LAN + VGA |
Built in QC Program | Yes: 3 Levels known, 3 Levels unknown |
Maintenance | Automatic Program |
Internal Patients Archive | Up to 6 months of patient data |
Power Supply | Universal Input 90/240V, 50-60Hz, 590W PFC |
Dimension | Height 27″ – Width 40″ – Depth 23″ |
Weight | 135 Kg – 298 lb |
Windows 2000®, Intel Pentium® and Cubitainer® are registered of Microsoft Corporation, Intel and Emergency Resources. IBM is registered trademark of the International Business Machine. | |