BT 3500

Automatic analyzer random access for clinical-chemistry and immunoturbidimetry computer controlled.

Enquiry

Description

Automatic analyzer random access for clinical-chemistry and immunoturbidimetry computer controlled. Automatic – upon programmation – pathological results repetition with or without automatic dilution. Automatic – upon programmation – pathological results in the panic range repetition. Urgent tests in actual time (STATs).

Sophisticated software for the system control. Software running under Windows 2000 Professional®. 40 on line tests mono or double reagent, refrigerated. 78 positions sample tray, programmable for samples, standards, controls and STATS.

Analyzer throughput: 360 test/h without ISE module. ISE module: 240 test/h. Two bar code readers for the positive identification of samples and reagents. Easy maintenance. No consumables.

Automatic washing of the reading cuvettes. Automatic – upon programmation – calibration of the I.S.E. module. Relation tests. Real pre-dilution up to 1:500. Clot Sensor: For Samples and Reagents.

Log file for data and performed operations.Vocal and on screen error messages.Touch screen monitor and wireless keyboard and mouse.

Reviews

There are no reviews yet.


Be the first to review “BT 3500”

TECHNICAL SPECIFICATIONS
TypeStand Alone, Multiparametric System
Analysis TypeClinical Chemistry, Turbidimetric, ISE
ThroughputUp to 360 Chem. Tests/hr + up to 240 ISE Tests/hr
ISE ModuleNa, K, Cl, CO² (indirect potentiometry)
Reagent Tray PositionsUp to 80 (40 large bottles + 40 small bottles)
Reagent Tray RefrigerationFrom 8°C to 15°C (programmable)
Sample Tray Positions78 (62 Samples + 16 Standard and Controls)
Sample tubesPrimary tubes and sample cups
Reaction volume180 up to 400 µl
Sample volume1.8 µl up to 100 µl
Reading cuvettes45 non-disposable optical glass
Sampling Arm2 for Serum and Reagents
Number of position78 Total – 52 Samples – 26 Standards and Control
Test ModesRandom Access, Batch, STAT
Test DirectoryUp to 500
Temperature controlPeltier System OFF/30/32/37°C (±0,2°C)
Vacuum PumpExternal
PhotometerSolid state 10 channel + Reference channel
Photometer Accuracy± 1% (0-2 O.D.), + 2,5% (2-3 O.D.)
Photometer Sensitivity± 0,001 Abs
Diluter1 – Long life ceramic 340 µl
Diluter Accuracy± 0,1% f.s.
Diluter Sensitivity± 0.1 µl
Clot SensorFor Sample and Reagents
Bar CodeFor Sample and Reagent
Remote DiagnosticsVia Modem (optional)
ComputerCPU Pentium M® 1.6Ghz or more,
Monitor color LCD 12″,
DVD Rom, HD=>80 Gb
Touch-screenYes: Glass Protected resistive kind
KeyboardStandard IBM® Compatible – wireless
MouseStandard – wireless
SoftwareWindows 2000® PRO
Process electronicµP. 1²CBUS Protocol
InterfaceRS232, Bidirectional + USB + LAN + VGA
Built in QC ProgramYes: 3 Levels known, 3 Levels unknown
MaintenanceAutomatic Program
Internal Patients ArchiveUp to 6 months of patient data
Power SupplyUniversal Input 90/240V, 50-60Hz, 590W PFC
DimensionHeight 27″ – Width 40″ – Depth 23″
Weight135 Kg – 298 lb
Windows 2000®, Intel Pentium® and Cubitainer® are registered of Microsoft Corporation, Intel and Emergency Resources.
IBM is registered trademark of the International Business Machine.
Features may change without notice